| Paramètre | Valeur |
|---|---|
| Article | Specifications |
| Poids | 450 g approx. |
| Détails | Controllers(Ethernet communications port) |
| Produit | Ultra High-speed- High-precision Laser Displacement Sensor |
| Matériel | Case: Polycarbonate |
| Accessoire | USB cable (2 m 6.562 ft long): 1 pc.Short bracket: 1 pc.Ferrite cores (E04SR200935A, by Seiwa Electric Mfg. Co., Ltd.): 3 pcs |
| Numéro de pièce | HL-C21C-P |
| Entrée de réinitialisation | PNP open-collector transistor- Reset is done when connected to external power (+).- Applied voltage: 30 V DC or less (Leak current: 0.1 mA or less) |
| Sortie d'alarme | PNP open-collector transistor- Maximum source current: 100 mA- Applied voltage: 30 V DC or less (between each output and +V)- Residual voltage: 1 V or less (at 100 mA source current) |
| Nom du produit | Ultra High-speed・High-precision Laser Displacement Sensor HL-C2 |
| Entrée de synchronisation | PNP open-collector transistor- ON at / during connection to external power (+)(depending on analysis mode)- Applied voltage: 30 V DC or less (Leak current: 0.1 mA or less) |
| Taux d'échantillonnage | 10 micro s, 20 micro s, 40 micro s, 100 micro s, 200 micro s, 400 micro s, 1 ms, 2 ms |
| Sortie stroboscopique | PNP open-collector transistor- Maximum source current: 100 mA- Applied voltage: 30 V DC or less (between each output and +V)- Residual voltage: 1 V or less (at 100 mA source current) |
| interface USB | USB 2.0 Full-speed (USB 1.1 compatible) compliant |
| Numéro de produit | HL-C21C-P |
| Tension d'alimentation | 24 V DC plus or minus 10 % including ripple 0.5 V (P-P) |
| Entrée zéro | PNP open-collector transistor- Zero set is ON when connected with external power (+).- Zero set turns to OFF after continuously connected to external power (+) for one second.- Applied voltage: 30 V DC or less (Leak current: 0.1 mA or less) |
| Indicateur : Puissance | Green LED (lights up at power on) |
| Consommation de courant | 500 mA approx. at 2 sensor heads connected350 mA approx. at 1 sensor head connected(100 mA approx. is additionally required when the mini console is connected) |
| Indicateur : Alarme 1 | Red LED (lights up when OUT1 can not be measured due to insufficient amount of light) |
| Indicateur : Alarme 2 | Red LED (lights up when OUT2 can not be measured due to insufficient amount of light) |
| Entrée de commande laser | PNP open-collector transistor- Laser emission is stopped when connected to external power (+).- Laser is emitted immediately after opened.- Applied voltage: 30 V DC or less (Leak current: 0.1 mA or less) |
| Entrée de modification de la mémoire | PNP open-collector transistor- Memory is specified when connected to external power (+).- Applied voltage: 30 V DC or less (Leak current: 0.1 mA or less) |
| Tête de capteur Connectale | Number of connectable units: Max. 2 units. |
| Entrée de verrouillage à distance | PNP open-collector transistor- Laser emission is delayed when connected to IL (+).- Laser emission stops at open- Applied voltage: 30 V DC or less (Leak current: 0.1 mA or less) |
| Affichage des paramètres / des données | Compact console (optional) |
| Sortie analogique : Courant | Current output scale: 4 to 20 mA / F.S (initial value)Output range during normal status: 2 to 24 mAOutput at abnormal status: 1 mA or 25 mAResolution: 3 micro A, Linearity plus or minus 0.05% F.S.Load impedance: 250 Ohm max.Response delay time: 10 micro s approx.(Note) The linearity is F.S.=16 mA to digital measurement value. Response delay time is the period after update of measurement value. |
| Sortie analogique : Tension | Voltage output scale: -5 to +5 V / F.S (initial value)Output range during normal status: -10.0 to +10.0 VOutput at abnormal status: -10.8 V or +10.8 VResolution: 2 mVLinearity: plus or minus 0.05 % F.S.Max. 2 mA, output impedance 50 OhmResponse delay time: 1.5 micro s / V approx.(Note) The linearity is F.S.=20 V to digital measurement value. Response delay time is the period after update of measurement value. |
| Résultat du jugement (HI, GO, LO) | PNP open-collector transistor- Maximum source current: 100 mA- Applied voltage: 30 V DC or less (between each output and +V)- Residual voltage: 1 V or less (at 100 mA source current) |
| Sortie d'alarme : Fonctionnement de sortie | Opened when the amount of light is insufficient |
| Sortie stroboscopique : Fonctionnement de sortie | Opened at data determination |
| Sortie d'alarme : Protection contre les courts-circuits | Incorporated |
| Sortie stroboscopique : Protection contre les courts-circuits | Incorporated |
| Indicateur : Tête de capteur A Rayonnement laser | Green LED (lights up during or immediately before laser emission of sensor head A) |
| Indicateur : Rayonnement laser de la tête de capteur B | Green LED (lights up during or immediately before laser emission of sensor head B) |
| Résistance environnementale : humidité ambiante | 35 to 85 %RH |
| Résistance environnementale : Résistance aux chocs | 196 m / s2 acceleration (20G approx.) in X, Y, and Z directions for three times each |
| Résistance environnementale : Température ambiante | 0 to +50 degrees +32 to +122 degrees Fahrenheit (No dew condensation or icing allowed), Storage: -20 to +70 degrees -4 to +158 degrees Fahrenheit |
| Sortie du jugement (HI, GO, LO) : Opération de sortie | Opened at output operation |
| Résistance environnementale : Résistance aux vibrations | 10 to 55 Hz frequency (period: 1 min.), 0.75 mm 0.030 in amplitude in X, Y and Z directions for 30 min. each |
| Sortie de jugement (HI, GO, LO) : Protection contre les courts-circuits | Incorporated |
Spécifications techniques et données de performance